Technical Settings

For the EO-150 ventilator, the AUTO trigger setting likely corresponds to an automatic patient-triggered cycle sensitivity that operates within a flow threshold range of 1.4L/min down to 0.7 L/min difference from the baseline flow (bias flow), allowing the ventilator to dynamically adjust the effort needed for the patient to initiate a breath.

The setting I Time Max = AUTO = 30 / Respiratory Rate set means the ventilator's Maximum Inspiratory Time is automatically calculated by taking the constant value of 30 seconds and dividing it by the set Respiratory Rate (breaths per minute), which serves as a back-up time limit for the inspiration phase to prevent the I time to be lasting more than half the set Respiratory Rate.

The Nominal Leak to AUTO applies a compensation of 40 l/min when the circuit is configured with a leak type.

BRAND MODEL LEAK @ 15 mbar
ALMS Primo F / Soft Facial / Soft Nasal / Soft Nasal Silent / Alnest N1 Silent 48 l/min
ALMS Alnest MC1 40 l/min
ALMS Primo P 38 l/min
ALMS Soft Baby / Soft Child 28 l/min
RESMED Quattro Air / Quattro / Quattro FX / Mirage / Mirage FX / Micro / Activa LT 48 l/min
RESMED Ultra Mirage II 40 l/min
RESMED Air Fit N10 / P10 45 l/min
RESMED Swift 2 / Swift FX 43 l/min
FISHER & PAYKEL Simplus / Free Motion RT040 41 l/min
FISHER & PAYKEL Eson 40 l/min
FISHER & PAYKEL PILAIRO 38 l/min
PHILIPS Amara 32 l/min
LOEWENSTEIN Joyce Full Face 46 l/min
LOEWENSTEIN Joyce One 32 l/min